Cam2 measure 10 probing1/13/2024 ![]() NOTE: The probe compensation stores information for the adapter. For more information, search for “ Sphere Compensation Method" in the User Manual for your FaroArm model. On the Probe Management screen, verify the TP20 Probe is active and then click the pencil icon to edit the probe values.Įnter the Probe Diameter and then enter 500-1000 for a De-bounce Time and then click Save.Ĭlick the back arrow (at the top right of the window) to return to Probe Management screen and click the Sphere Compensation button.Ĭlick Start Compensation and follow the on-screen instructions to compensate the probe. Press the “P” key or on the Home tab, click the Device Center icon to open the Device Center window. Verify the FaroArm you are using is selected on the right and then click Probe Management. Open the FaroArm manager, Start> FARO> FaroArm Manager If your FaroArm does not have the option port, it must be returned for service to have the port added (part# 1001-1024).Īfter the touch probe/adapter is installed… To use the TP20 probe, this option must have been selected at the time of order. Once complete then reattach the TP20 touch probe. However, you must compensate the 8-Axis using the 6mm or 3mm hard probe. TP20 touch probes are supported on the Quantum/Quantum Max FaroArm with the 8-Axis. † TP20 touch probes are not supported on the Edge FaroArm with the 8-Axis. ![]() 3D Hand Held Scanner - Freestyle3D Objects ![]()